繁體中文   |   English
 

Real-time PCR
StepOne
 
Product Description:
Thermal Cycling System Peltier-based system
Block Format • 48-well block
Compatible Consumables • 48-well plates with optical flat caps
• 48-well plates with optical adhesive
   cover
• 8-tube (0.1 mL) strips with optical flat
   caps
• Individual (0.1 mL) tubes with optical
   flat caps
Supported Volumes 10–30 μL
Sample Ramp Rate • Fast mode: +/- 2.2 °C/sec
• Standard mode: +/- 1.6°C/sec
Peak Block Ramp Rate 4.6 °C/sec
Temperature Range 4°C-100°C
Temperature Accuracy  +/-0.25°C (35°C to 95°C) of setpoint/display temperature, measured at 3 minutes after clock start.
Temperature Uniformity  +/-0.50°C, 30 seconds after clock start.
Melt Curve Resolution As small as 0.1°C
Optical System LED, emission filters, and photodiode
Calibrated Dyes at Installation SYBR® Green I, FAM™, VIC®, JOE™, ROX™ dyes
Passive Reference Dyes ROX™ or any calibrated dye. Option exists to select no passive reference.
Data Collection Data collected in all 3 filters for all wells regardless of plate setup.
Plate setup may be altered after run completes.
Fast cycling: Less than 40 minutes
Quantitative PCR Run Time Standard cycling: Less than 2 hrs
Fast cycling: Less than 40 minutes
Touchscreen LCD/6.5 in. Full VGA (640x480)/32K colors
Instrument Dimensions • Width: 24.6 cm (9.7 in.)
• Depth: 42.7 cm (16.8 in.)
• Height: 51.2 cm (20.2 in.)
• Weight: 23.6 kg (52 lbs.)
Electrical US/Canada:
• Voltage: 120 VAC ± 10%
• Frequency: 50/60 Hz ± 1%
• Nominal Current Draw: 8A
• Power: 950 W

StepOnePlus
 
Product Description:
VeriFlex™ Block 25°C (5°C zone-to-zone)
Thermal Cycling System Peltier-based system
Block Format 96-well block
Compatible Consumables • 96-well (0.1 mL) plates with optical
   flat caps
• 96-well (0.1 mL) plates with optical
   adhesive cover
• 8-tube (0.1 mL) strips with optical flat
   caps
• Individual (0.1 mL) tubes with optical    flat caps
Supported Volumes 10–30 μL
Sample Ramp Rate • Fast mode: +/- 2.2 °C/sec
• Standard mode: +/- 1.6°C/sec
Peak Block Ramp Rate 4.6 °C/sec
Temperature Range 4°C-100°C
Temperature Accuracy  +/-0.25°C (35°C to 95°C) of setpoint/display temperature, measured at 3 minutes after clock start.
Temperature Uniformity  +/-0.50°C, 30 seconds after clock start.
Melt Curve Resolution As small as 0.1°C
Optical System Blue LED excitation source, four-emission filters, and photodiode
Calibrated Dyes at Installation FAM™, SYBR® Green I, VIC®, JOE™, NED, TAMRA, ROX™ dyes
Passive Reference Dyes ROX™ or any calibrated dye. Option exists to select no passive reference.
Data Collection Data collected in all 3 filters for all wells regardless of plate setup.
Quantitative PCR Run Time Standard cycling: Less than 2 hrs
Fast cycling: Less than 40 minutes
Touchscreen LCD/6.5 in. Full VGA (640x480)/32K colors
Instrument Dimensions • Width: 24.6 cm (9.7 in.)
• Depth: 51.2 cm (20.2 in.)
• Height: 51.2 cm (20.2 in.)
• Weight: 24 kg (53 lbs.)
Electrical US/Canada:
• Voltage: 120 VAC ± 10%
• Frequency: 50/60 Hz ± 1%
• Nominal Current Draw: 8A
• Power: 950 W

7300
 
Product Description:
Thermal Cycling System Peltier-based system
Block Format 96-well block
Compatible Consumables 96-well plates and 0.2 mL tubes
Supported Volumes 20 – 100 μL
Sample Ramp Rate +/- 1.1°C/sec
Peak Block Ramp Rate 1.6°C/sec
Temperature Range 4°C-100°C
Temperature Accuracy  +/-0.25°C (35°C to 95°C) of setpoint/display temperature, measured at 3 minutes after clock start.
Temperature Uniformity  +/-0.50°C, 30 seconds after clock start.
Optical System Tungsten-halogen lamp excitation source.  
Single-excitation filter, four-emission filters, and CCD camera
Calibrated Dyes at Installation SYBR® Green I, FAM™, VIC™, JOE™, NED™, TAMRA™, ROX™
Passive Reference Dyes ROX™ or any calibrated dye. Option exists to select no passive reference.
Additional Dyes Available Calibration for new dyes within the wavelength range is possible by following the custom dye calibration procedure in the User's Manual. Purchase of additional filter sets is not necessary.
Data Collection Data collected in all 4 filters for all wells regardless of plate setup.
Plate setup may be altered after run completes.
Quantitative PCR Run Time Less than 2 hrs
Electrical US/Canada:
• Voltage: 120 VAC ± 10%
• Frequency: 50/60 Hz ± 1%
• Nominal Current Draw: 8A
• Power: 950 W

7500
 
Product Description:
Thermal Cycling System Peltier-based system
Block Format 96-well block
Compatible Consumables 96-well plates and 0.2 mL tubes
Supported Volumes 20 – 100 μL
Sample Ramp Rate • Standard Mode: +/-1.6°C/sec
• 9600 Emulation Mode: +0.8 and
    -1.6°C/sec
Peak Block Ramp Rate 2.5°C/sec
Temperature Range 4°C-100°C
Temperature Accuracy  +/-0.25°C (35°C to 95°C) of setpoint/display temperature, measured at 3 minutes after clock start.
Temperature Uniformity  +/-0.50°C, 30 seconds after clock start.
Optical System Tungsten-halogen lamp excitation source.  
Five-excitation filters, five-emission filters and CCD camera
Calibrated Dyes at Installation SYBR® Green I, FAM™, VIC™, JOE™, NED™, TAMRA™, ROX™, Texas Red®, Cy3™, Cy5™
Passive Reference Dyes ROX™ or any calibrated dye. Option exists to select no passive reference.
Additional Dyes Available Calibration for new dyes within the wavelength range is possible by following the custom dye calibration procedure in the User's Manual. Purchase of additional filter sets is not necessary.
Data Collection Collects data for all 5 filters for all wells regardless of plate setup.
Plate setup may be altered after run completes.
Quantitative PCR Run Time Less than 2 hrs
Electrical US/Canada:
• Voltage: 120 VAC ± 10%
• Frequency: 50/60 Hz ± 1%
• Nominal Current Draw: 8A
• Power: 950 W

7900HT
 
Product Description:
Thermal Cycling System Peltier-based system
Block Format • 96-well block (Standard)
• 96-well block (Fast)
• 384-well block
• Low Density Array block
Compatible Consumables • 96-well plates (Standard)
• 96-well plates (Fast)
• 384-well plates
• Low Density Arrays
Supported Volumes • 96-well Standard: 25-100 μL
• 96-well Fast: 10-30 μL
• 384-well: 5-20 μL
Sample Ramp Rate • Fast Mode:  Ave. +/- 3.0°C/sec
• Standard Mode: +/-1.6°C/sec
• 9600 Emulation Mode: +0.8 and -1.6°C/sec
Peak Block Ramp Rate • 96-well (Fast): 4.8°C/sec
• 96-well (Standard), 384-well, Low Density Array:
2.0°C/sec
Temperature Range 4°C-100°C
Temperature Accuracy  +/-0.25°C (35°C to 95°C) of setpoint/display temperature, measured at 3 minutes after clock start.
Temperature Uniformity  +/-0.50°C, 30 seconds after clock start.
Optical System Extended-life 488 nm argon-ion laser excitation source.  
Excitation light distributed to all wells via a  
dual-axis synchronous scanning head.  
Fluorescence detection via spectrograph and CCD camera
Calibrated Dyes at
Installation
Low Density Array: FAM™, VIC™, ROX™
All other blocks: SYBR® Green Dye I, FAM™, VIC™, JOE™, NED™, ROX™, TAMRA™, TET™
Passive Reference Dyes ROX™ or any calibrated dye. Option exists to select no passive reference.
Data Collection Collects data for all detectable fluorescence within the spectrograph range.
Plate setup may be altered after run completes.
Quantitative PCR Run Time 33 minutes (96–well in fast mode), 52 minutes (384-well in fast mode),
<2 hrs (96 and 384-wells in standard/9600 emulation mode and Low Density Array)
Electrical • Voltage: 200-240 VAC ± 10%
• Frequency: 50-60 Hz ± 1%
• Maximum Current Draw: 16A